Type: |
Range: |
The oxygen sensor consists of an oxygen sensitive
electrode with a processing unit (Oxygen Adapter). The central part of the electrode
is a chamber separated from the probe (gas or liquid) by a semi-permeable membrane.
In the vicinity of this mem-brane the partial pressure of oxygen is measured.
The Oxygen Adapter is used to set the electrode up for measurement in gas or liquid,
or to the calibration mode.
The oxygen electrode is based on the Clarke cell principle using a
polarographic measuring technique. The sensing element consists of a silver
anode and a gold cathode. The silver anode is held at a potential of +800mV
(polarizing voltage) with respect to the gold cathode. Cathode and anode are
immersed in an electrolyte and separated from the sample by a membrane permeable
to oxygen.
Oxygen is reduced at the cathode in the following reaction
creating a point of zero Oxygen tension (partial pressure) at the
cathode:
Provided that diffusion through the membrane is the
rate-controlling step in the path of the Oxygen to the cathode, the signal
current flowing from the anode to the cathode is proportional to the
concentration (partial pressure) of oxygen in the sample.
The adapter
converts this current signal into voltage, compensates for the sample
temperature, allows the output to be calibrated and scales it appropriately for
the MultiLog.
Temperature Compensation A 10 KOhm thermistor is used to
compensate the output for changes in the Oxygen permeability of the membrane
with temperature. The thermistor is located under the metal disk on the side of
the electrode about 35mm above its lower end and should be immersed in the
sample medium and at the same temperature as the membrane.
You perform the calibration directly from the Oxygen Adapter. Note that you have to calibrate your sensor always for aqueous solution. Define the sensor as described above and start recording to monitor the procedure of calibration online.
Water/aqueous solution (100% DO2): Select 'DO2 sol' from the Control Panel. Stir a sample of distilled water in an Erlenmeyer flask vigorously for several minutes to make sure that the air above is saturated with water vapor. Position the electrode approximately 1cm above the sample. Start logging and adjust the CAL knob until you see the MultiLog showing 100%. Air (25% O2): Select 'O2 air' from the Control Panel. Start logging and adjust the CAL knob until you see the MultiLog showing 20%.
Note: No compensation is provided for changing atmospheric pressure. Calibration should therefore be carried out at the pressure at which measurements are to be made.
| The figure in the right shows a typi-cal graph received from following the oxygen formation in a photosyn-thesis process of a water plant. | ![]() |
Treat the membrane carefully as it is fragile. No grease, oil or organics should come into contact with the membrane. Avoid contact with the membrane to prevent punctures.
Range: 0%-125% Notice that the DB-Lab Software will cut everything above 125%, while the MultiLog will still show more. This is done, because the adapter is inaccurate above 125% DO2.
Resolution: +/- 0.25% DO2
Electrode