The DT116 is
equipped with built in GM-tube and integral power supply. This will allow direct
connection to the input of the MultiLog that will also provide the 5V required
operation voltage. The DT116 is equipped with a power indicator, indicating that
an operational voltage is supplied to the unit- a yellow LED. Furthermore, the
unit is also equipped with a buzzer emitting an audible signal when each pulse
is recorded. The DB116 is sensitive to Alpha, Beta and Gama radiation.
The Tube is sensitive to Alpha, Beta and Gamma radiation
| Plateau threshold voltage (Vb1) | 450 V |
| Plateau Length (Vb2-Vb1) | 150 V |
| Recommended supply voltage | 500 V |
| Thickness | 1.5 to 2.0 mg/cm2 |
| Useful Diameter | 9.14 mm |
| Material | Mica |
| Gas Filling | Neon, Argon and Halogen |
| Dead Time (at recommended supply Voltage) | 90ms |

The Plateau is the
part of the diagram where the number of counts per second is (almost)
independent of the voltage The graph in the right expresses the radioactivity
magnitude versus the distance between the radioactive source and the Geiger
Muller sensor.